The
TX2000 is the unique spectrometer that allows to perform
Energy Dispersive X-Ray Fluorescence Analysis in both
Total Reflection and Traditional (45 degrees) Geometry. |
| |
|
Principal
advantages of TXRF
|
- No
matrix effects
- A
single internal standard greatly
simplifies quantitative analyses
- Calibration
and quantification independent from any
sample matrix
- Simultaneous
multi-element ultra-trace analysis
- Several
different sample types and applications
- Minimal
quantity of sample required for the
measurement (5 ml)
- Unique
microanalytical applications for liquid
and solid samples
- Excellent
detection limits (ppt or pg) for all
elements from sodium to plutonium
- Excellent
dynamic range from ppt to percent
- Possibility
to analyse the sample directly without
chemical pretreatment
- No
memory effects
- Non
destructive analysis
- Low
running cost
|
|
|
|
|
|
|
|
|
Principles of the TXRF analytical method |
|
TXRF
is founded on the same principles of the EDXRF with,
however, one significant difference. In contrast to
EDXRF, where the primary beam strikes the sample at an
angle of 45°, TXRF uses a glancing angle of a few
milliradians. Owing to this grazing incidence, the
primary beam is totally reflected. By illuminating the
sample with a beam that is being totally reflected,
absorption of the beam in the supporting substrate is
largely avoided and the associated scattering is greatly
reduced. This also reduced the background noise
substantially. A further contribution to the reduction
of the background noise is obtained by minimising the
thickness of the sample. A small drop of the sample
(5-100 microliters of the substance dissolved in an
appropriate solvent) is placed on a silica carrier. On
evaporation of the solvent a thin film, a few nanometers
thick, remains. In practice the greater part of the
scattering normally arising from the sample and its
matrix is eliminated. This is because matrix effect
cannot build up within minute residues or thin layers of
a sample. Besides its high detection power, simplified
quantitative analysis in made possible by internal
standard. |
|
|
| |
|
| |
|
Applications |
- Environmental
Analysis: water, dust, sediments, aerosol
- Medicine:
toxic elements in biological fluids and tissue
samples
- Forensic
Science: analysis of extremely small sample
quantities
- Pure
chemicals: acids, bases, salts, solvents, water,
ultrapure reagents
- Oils
and greases: crude oil, essential oil, fuel oil
- Pigments:
ink, oil paints, powder
- Semiconductor
Industry: by VPD (vapor-phase decomposition)
- Nuclear
Industry: measurements of radioactive elements
|
|
|
|
|
|
|
The
main features of the TX 2000 spectrometer |
Hardware:
- TXRF
and EDXRF (traditional 45° geometry) spectroscopy
in the same equipment
- Automatic
switching of primary beam
(MoKa,
WLa/Lb
and
bremsstrahlung
33 keV) using double
anode Mo/W X-ray tube, based on innovative software.
We select the energy required using an high
reflectivity 80% (WLa/Lb/MoKa)
multilayer.
We can choose also other X-ray tubes and
monochromatise the energy that you need
- 3.8
liters UHV (Si(Li) 20 mm2 detector area)
high resolution detector <133 eV (Ka
Mn radiation at 5.89 keV), with an ultra-thin and
highly corrosion resistant window (8 mm
Dura-Beryllium)
- Minimal
distance between the sample and the detector
(mounted to the axis normal to the plane of the
sample.) In this position the detector is also
completely out of the primary beam, as the angle
between the incident and the reflected beams is so
large
- Instrumental
detection limits for more than 50 elements below 10
pg.
- Helium
device to improve the detection limits for the light
elements
- The
Spectrometer is fully automated and you can control
different total reflection conditions for different
energies from the PC, using stepping-motors mowing
monochromator and tube shield and MS Windows
software
|
Software:
- TXRFACQ32
acquisition program, 32 bit version for windows
98/ME/2000 which allows one to accomplish the
following functions: X-ray generator load settings,
multisample positioning, counter chain parameter
settings, selection of radiation, centring
procedure, K, L, & M markers, time or count
selection, acquisition of data in both geometries (TXRF
and EDXRF).
- EDXRF32 for the elaboration of data 32 bit version for Windows 98/ME/2000, includes:
Least square Marquardt fit procedure for the area calculation (spectral analysis), automatic/manual search function, manual or automatic calibration of energy, quantification via an internal standard using theoretical and experimental sensitivity curves for total reflection, several types of background correction.
For the 45 degrees geometry the following methods are available:
Emission/Transmission (for powder/light matrix samples).
Fundamental parameters using relationship between incoherent and coherent radiations for evaluation of the dark matrix.
Fundamental parameters (for alloys).
Thin film.
Semi-empirical corrections for matrix effect.
A set of tools for calculating: fundamental parameters, fluorescence yield, edge jump, detector efficiency, differential cross section (coherent and incoherent), atomic form factor, incoherent scattering function, elements lines and ratios, stoichiometry
|
|
|
|
|
|
|
Example
of detection limits - Chromium in distilled water |
|
Tungsten
Radiation (40 kVolt – 30 mA)
|
|
|
| Concentration
of Cr (ppb) |
Volume
ml (5 x N) |
Live
Time (seconds) |
Detection
Limit (ppt) |
Detection
Limit (pg)= ppt x ml/1000 |
| 24.5 |
10
(5 x 2) |
500 |
370 |
3.70 |
| 24.5 |
50
(5 x 10) |
500 |
120 |
6.00 |
| 24.5 |
50
(5 x 10) |
300 |
170 |
8.50 |
| *
24.5 |
100
(5 x 20) |
500 |
70 |
7.00 |
| 24.5 |
100
(5 x 20) |
1000 |
55 |
5.50 |
| 24.5 |
100
(5 x 20) |
5000 |
35 |
3.50 |
| 1.97 |
10
(5 x 2) |
500 |
400 |
4.00 |
| 1.97 |
10
(5 x 2) |
300 |
440 |
4.40 |
| 1.97 |
50
(5 x 10) |
500 |
80 |
4.00 |
| 1.97 |
50
(5 x 10) |
300 |
125 |
6.25 |
|
|
|
|
* See
the attached spectrum |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| |
|
| |
|
| |
Some TXRF
spectra, from different fields of analysis
|
| |
|
|
|
|
|
|
If
you are interested, we can analyse your samples free of charge, so
that you can see the potentiality of this new generation of
instruments.
|
|
|
|